manualshive.com logo in svg
background image

P/N: 10-1000 

 Revision: F, Issued: October 2019 

 Page: i

  

 

 

 

 

 

Thank you for purchasing the OdaLog

®

 portable gas logger manufactured by 

Thermo Fisher Scientific Australia Pty Ltd. 

In our quest to continually improve the product we manufacture, we look 
forward to your feedback on your experience using the OdaLog

®

 in different 

environments and applications, as well as the usefulness of the information 
you collect using it’s data-logging function. 

If you have any suggestions or comments, please contact your distributor or 
alternatively e-mail us direct on

 

[email protected]

 

 

 

Gas Logger 

 

 
 
 

 

COPYRIGHT© 2018 

The sole purpose of this document is to assist in the operation of the OdaLog

®

 portable gas logger. 

We do not convey any right to use this information for any purpose other than in the operation or maintenance 
of the OdaLog

®

 portable gas logger. 

OdaStat™ & OdaStat-G™ © Software Copyright 2008 (Thermo Fisher Scientific Australia Pty Ltd) 

 

GENERAL DISCLAIMER 

This Manual summarises our knowledge of the OdaLog

®

 portable gas logger including how to use it based on 

the information available at the time.  You should read this Manual carefully and consider the information in the 
context of how the instrument is to be used.   

The OdaLog

®

 portable gas logger is sold subject to our standard terms and conditions of sale and any written 

warranty given by Thermo Fisher Scientific.  Thermo Fisher Scientific does not accept any other liability either 
directly or indirectly for any losses suffered in connection with the use and application of the OdaLog

®

 portable 

gas logger, whether or not in accordance with any advice, specification, recommendation or information given 
by it or as a result of any errors or omissions in this Manual. 

Содержание OdaLog LL

Страница 1: ...se other than in the operation or maintenance of the OdaLog portable gas logger OdaStat OdaStat G Software Copyright 2008 Thermo Fisher Scientific Australia Pty Ltd GENERAL DISCLAIMER This Manual summ...

Страница 2: ...CTION MANUAL LOGGING MODE ONLY DEFAULT ENABLED 10 6 6 STOP LOGGING FUNCTION MANUAL LOGGING MODE ONLY DEFAULT ENABLED 10 6 7 AUTO LOG START FUNCTION DEFAULT DISABLED 11 6 8 IRDA INFRA RED DATA COMMUNIC...

Страница 3: ...PARTS 30 10 3 GAS TEST KITS 31 10 4 REPLACEMENT TEST GAS CYLINDERS 31 11 WARRANTY 32 11 1 WARRANTY 32 11 2 WHAT THE WARRANTY DOES NOT COVER 32 11 3 WHAT THERMO FISHER SCIENTIFIC WILL DO 32 11 4 LIMIT...

Страница 4: ...GROUP IIC THIS CONDITION DOES NOT APPLY FOR GROUPS IIA AND IIB Table 1 Approved Battery Types Manufactures Part Number Code Type Size Saft LS26500 Lithium C Tadiran TL 5920 S Lithium C 1 3 Electromagn...

Страница 5: ...SIGNED TO OPERATE AS A SAFETY DEVICE WHEN ENTERING CONFINED SPACES AND TOXIC HAZARD ENVIRONMENTS ALL APPROPRIATE REGULATIONS AND OCCUPATIONAL HEALTH AND SAFETY PRECAUTIONS SHOULD BE STRICTLY ADHERED T...

Страница 6: ...ogged data without the need for any disassembly The extra large easy to read LCD makes the OdaLog simple to use and interpret instrument data and settings on site The instrument main features can be v...

Страница 7: ...OdaLog Gas Logger User Manual P N 10 1000 Revision F Issued October 2019 Page 4...

Страница 8: ...of the OdaLog instrument with sensors fitted should be limited to the temperature range of 0 C 32 F to 25 C 75 F and between 20 to 80 RH The following is a list of gas types available for the OdaLog...

Страница 9: ...lphabetic characters some numerical values displayed on some screens 3 Battery Level Indicator Displays the battery levels 4 stage depletion see Section 6 12 4 Communications Symbol Indicates if the i...

Страница 10: ...using supplied calibration fitting Part No 11 0001 at the flow rate and for the flow time as detailed in Table 5 or until reading stabilises and record End Level result when reading stabilises See Sec...

Страница 11: ...OdaLog Gas Logger User Manual P N 10 1000 Revision F Issued October 2019 Page 8 6 OPERATION 6 1 Operation Overview...

Страница 12: ...the button during the countdown routine will cause exit to OFF screen Note 2 Peak is constantly updated when OdaLog is in use The Peak value is reset at switch on and at the start of each logging sess...

Страница 13: ...ay and the LED s on either side of the display flashing once every 10 seconds Default Enabled can be changed in the OdaStat G software Note 1 If the Logger memory is full and you have Stop log when fu...

Страница 14: ...e they are functioning correctly 6 8 2 Exit IrDA mode Infra Red Data Communication Mode Pressing the button when in IrDA mode or waiting 75 seconds for unit to time out will cause the instrument to ex...

Страница 15: ...R device to your PC and install the OdaStat G software package Open the OdaStat G software package on your PC and ensure communication setting are configured correctly to communicate with your OdaLog...

Страница 16: ...mmunication mode 6 9 3 Pairing Instrument with PC The Odalog instrument must be paired with the PC before the communication can be established Pairing procedures Enter BTPAIR mode on the instrument LC...

Страница 17: ...ss Auto Log Start is active See Section 6 7 Note 2 Releasing the button during the countdown routine will cause exit to instantaneous screen 6 11 Clear Logged Data Switch instrument OFF Press hold but...

Страница 18: ...the word FLAT continuously on the display with the Battery Warning symbol flashing Note The instrument cannot be turned back on until the battery has been replaced Refer Section 8 1 1 for battery repl...

Страница 19: ...fault Enabled This function is activated via the OdaStat G software package Instrument Parameters section by selecting the Suppress Negative Reading option Once active the instrument suppresses all ne...

Страница 20: ...rs listed are likely to respond to 100 ppm of various interfering gases There may be gases not on this chart which also interfere with OdaLog sensors Table 4 Sensor Cross Sensitivity Approximate respo...

Страница 21: ...onments See Section 9 8 7 2 3 Pressure Sensors may exhibit effects to changes in pressure For example an increase in pressure may result in an increase in gas reading For this reason we do not recomme...

Страница 22: ...a replacement part Only use Thermo Fisher Scientific replacement parts The OdaLog should be switched off when not in use and stored in a clean and dry environment as usually found in a normal office...

Страница 23: ...ew battery that the positive terminal insulator is not damaged this insulator may cause the battery to break contact if the insulator tab damaged Figure 3 Removing Battery and Replacing Battery Smear...

Страница 24: ...in the OdaStat G software package Note Some sensors will require up to 24hrs to settle after battery replacement Figure 4 Top Cap Alignment 8 1 2 Filter Replacement Figure 5 Filter Replacement Ensure...

Страница 25: ...p retaining screws DO NOT OVERTIGHTEN Note 1 Filter orientation is important Ensure that the replacement filter is correctly placed with the shiny side facing outward Note 2 Some sensor filters are fi...

Страница 26: ...ixtures than others particularly hydrogen sulphide Advice in relation to suitability of regulators tubing should be sought from the calibration mixture supplier 9 3 Tools and Test Equipment The follow...

Страница 27: ...Minimum Flow Rate ml min Gas flow time Hydrogen sulphide 50 ppm H2S 250 500 5 Minutes H2S 250 500 15 Minutes Carbon monoxide 500 ppm CO 250 500 3 Minutes Sulphur dioxide 20 ppm SO2 500 5 Minutes Nitr...

Страница 28: ...trument is in Null mode Place magnetic base of Allen Tool Part No 11 0000 provided on the left hand side of the LCD display above the wording NULL DONE will appear briefly and the LCD will return to t...

Страница 29: ...After waiting for the time shown in Table 5 the gas reading on screen should have stabilised and calibration can be performed as follows Ensure gas is still flowing over sensor Press hold down button...

Страница 30: ...rning off gas regulator and removing calibration fitting Ensure that reading returns to close to zero within 5 minutes in clean air and re zero if necessary Note The instrument will TIMEOUT and return...

Страница 31: ...commended flow rate and time or until reading stabilises as detailed in Table 5 Record reading This will be the Start Level note reading in relation to level stated on calibration mixture cylinder 9 8...

Страница 32: ...ration gas at the end of a logging period N B The above decay compensation assumes gas exposure to the sensor has occurred in a uniform way rather than in large bursts Accordingly we have used a strai...

Страница 33: ...ent seals o rings new filter s grill button cover moisture absorbtion pack s 22 0002 Button Cover Moulded Pk 10 02 0136 Lithium Battery Tadiran TL 5920 S 02 0137 Lithium Battery Saft LS26500 13 0107 2...

Страница 34: ...itrogen dioxide 10ppm Hydrogen cyanide 10ppm Hydrogen chloride 10ppm Ammonia 50ppm Chlorine 5ppm Hydrogen 200ppm Ethylene Oxide 5ppm Nitrogen 100 VOL Oxygen in Nitrogen 20 9 VOL 10 4 Replacement Test...

Страница 35: ...ed costs including without limitation costs incurred in replacing filters or gaskets This warranty is subject to Thermo Fisher Scientific s terms and conditions of sale 11 4 Limitations The obligation...

Страница 36: ...arge FAILED Sensor failure during Span operation INST Instantaneous measurement IrDA Infra Red Data Communication Mode MEMORY Logging memory used Status Menu Start logging Main Menu MEMORY FULL Data l...

Страница 37: ...OdaLog Gas Logger User Manual P N 10 1000 Revision F Issued October 2019 Page 34 14 Appendix B CERTIFICATES...

Страница 38: ...OdaLog Gas Logger User Manual P N 10 1000 Revision F Issued October 2019 Page 35...

Отзывы: