manualshive.com logo in svg
background image

Operating manual

OPUS20E 
for external sensors

BA-OP20E-03-EN

Содержание OPUS20E

Страница 1: ...Operating manual OPUS20E for external sensors BA OP20E 03 EN ...

Страница 2: ...onclusions and deriving actions is the user The correctness of the results presented is excluded from any liability or guarantee Liability for damages which have been caused by utilising the presented measured results is strictly excluded 2 Intended use The OPUS20E data logger is designed to detect and record a range of measured values that are detected by the measuring device and the connected ex...

Страница 3: ...ecording Display for inactive measured value recording Display for date Display for time Mode selection button Data logger for connecting to external sensors 3 Equipment 5 pin round connector M12 for connecting up to four digital BUS compati ble sensors 10 pin terminal with two independent input channels for connecting up to two analogue sensors Pin 2 Data B Pin 1 Terminal 24V input ext supply for...

Страница 4: ...s left and the display returns back to the original mode without making changes This function is only available if a BUS sensor has already been registered M61 M81 An 0 M1 M2 M3 M4 M51 M52 M53 M6x M81 An 1 M7x 4 Scope of supply The following components are included in the standard scope of supply Data logger USB connection cable CD ROM with operating manual SmartGraph software and software manual ...

Страница 5: ...ecessary that UDP broadcasts are allowed through the network b Note Based on their concept UDP broadcasts do not operate through a router NAT but instead only within its own network If the data logger is outside of the network using fixed IP addresses through a system administrator is highly recommended Connect to the network If the data logger is connected to a local network and the network funct...

Страница 6: ...he BUS sensor to the 5 pin round connector of the data logger see chapter 3 The sensor is supplied with current via the plug By optionally using up to three Y distributors a maxi mum of four digital BUS sensors can be connected to the data logger Further information about the currently available BUS sensors is provided in the technical data in chapter 11 7 1 2 Configuration of BUS sensors M6x So t...

Страница 7: ...l 2 has pins 2 4 6 8 and 10 see chapter 3 Both of the input channels support the sensor configurations listed in the following subchapters Example display Registration function M61 selection for registering the first BUS sensor n1 Example display Message specifying the BUS ID 71 01 for the first registered BUS sensor Example display Log off function M70 selection for logging off the first selected...

Страница 8: ... for the first input channel Alternatively navigate to the configuration mode M81 AN1 to configure the specification of a sensor for the second input channel The option for configuring the first analogue sensor is shown on the display D0 D1 B0 B1 A0 A1 Pt100 D0 D1 B0 B1 A0 A1 C0 C1 Pt100 TC TC B0 B1 A0 A1 Sensor At initial startup there is no sensor assignment preset in the factory at later config...

Страница 9: ...rther detailed information about configuration of the processing channels in the software is provided in the software manual which you can open by us ing the help function of the SmartGraph software 8 3 Displaying measured values One of the channel groups specified in table 1 for displaying measured values can be configured to be displayed on each of the three measured value rows Here the current ...

Страница 10: ...tion 0 20 mA Current mA cur min max mid PT100 temperature measurement in 3 cable operation Temperature C cur min max mid PT100 temperature measurement in 3 cable operation Temperature F cur min max mid PT100 temperature measurement in 4 cable operation Temperature C cur min max mid PT100 temperature measurement in 4 cable operation Temperature F cur min max mid Temperature measurement with thermal...

Страница 11: ...display Additionally a hysteresis can be set by which the measured value must return to the valid range to switch off the alarm Example display for an alarm result of the measuring channel shown in the first measured value row Alarm display Upper alarm limit value Valid value corridor good range Lower alarm limit value flashing static Logging mode Reading the measured value memory 9 3 Acoustic alar...

Страница 12: ...he device is not possible and it does not respond to requests from the Smart Graph3 software After memory has been completely deleted FOMA FLSH 100 is shown on the display Afterwards the display returns to normal operation Further detailed information about using the software is provided in the software manual which you can open from the help function of the Smart Graph software Display for deleti...

Страница 13: ... 500 C Accuracy 0 2 C 0 1 of the measured value Resolution 0 02 C Memory organisation Sensing interval 10 30 s 1 10 12 15 30 min 1 3 6 12 24 h Saving interval 1 10 12 15 30 min 1 3 6 12 24 h Data storage 16 MB approx 3 200 000 measured values Data recording up to 20 simultaneous measuring channels Configuration and dimensions LCD display W 90 x H 64 mm Housing Plastic Dimensions L 180 x W 32 x H 7...

Страница 14: ...egative polarity approx 0 024 V 0x53 CHANNEL_UNDERRANGE Limit range 0 024 V 0 000 V Displayed 0 000V Normal operation mode 0 000 V 1 000 V Limit range 1 000 V 1 024 V Displayed 1 000V Overvoltage approx 1 024 V 0x52 CHANNEL_OVERRANGE Special case open Input floating e g CHANNEL_OVERRANGE so 0x52 4 20 mA Negative polarity approx 0 48 mA 0x53 CHANNEL_UNDERRANGE Sensor breakage approx 0 48 mA 0 30 mA...

Отзывы: