Command Reference
R&S
®
ZNA
754
User Manual 1178.6462.02 ─ 12
.......................................................................................... 998
............................................................................................1039
........................................................................................1044
..........................................................................................1047
...........................................................................................1215
............................................................................................ 1279
...........................................................................................1282
.............................................................................................1315
...........................................................................................1318
.............................................................................1327
8.3.1
CALCulate Commands
The
CALCulate...
commands perform post-acquisition data processing. Functions
in the
SENSe
subsystem are related to data acquisition, while the
CALCulate
subsys-
tem operates on the data acquired by a
SENSe
function.
8.3.1.1
CALCulate:CLIMits...
The
CALCulate:CLIMits...
commands control the composite limit check.
CALCulate:CLIMits:FAIL?
Returns a 0 or 1 to indicate whether or not a global, composite limit check on several
traces has failed.
Since V2.20 of the R&S
ZNA FW the result is automatically recalculated whenever a
relevant setting is changed, i.e. a subsequent query will return the updated limit viola-
tion state.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:CLIM:FAIL?
Query the result for the composite limit check. As only one trace
is tested, the response should be equal to the previous
response.
Usage:
Query only
Manual operation:
See
8.3.1.2
CALCulate:DATA...
The
CALCulate:DATA...
commands provide access to the results of a measure-
ment.
SCPI Command Reference