Command Reference
R&S
®
ZNA
1092
User Manual 1178.6462.02 ─ 12
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Parameters:
<PowerMeter>
Number of external power meter. The parameters UP, DOWN,
MIN, MAX are not available for this command.
Range:
1 to number of configured external generators
*RST:
The power meter selection is not changed by a
reset of the analyzer.
<TestPort>
Test port number. During power calibration, the power meter is
connected to this port.
*RST:
1 to the number of test ports
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
Manual operation:
See
[SENSe:]CORRection:IMEThod
<InterpolationMethod>
Selects the algorithm for the interpolation of user system error corrections.
For the factory system error correction data, the analyzer always uses linear interpola-
tion.
Parameters:
<InterpolationMethod>LINear | HORDer
LINear
Linear interpolation
HORDer
Higher order (cubic spline) interpolation
Manual operation:
See
[SENSe<Ch>:]CORRection[:STATe]
<Boolean>
Enables or disables the system error correction for channel
<Ch>
.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the correction.
*RST:
ON
Example:
*RST; :CORR?
Reset the instrument and query whether channel 1 is system
error corrected. The response is
1
.
Manual operation:
See
SCPI Command Reference