Command Reference
R&S
®
ZNA
1220
User Manual 1178.6462.02 ─ 12
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:ARBitrary:EFRequency<Gen>
<Boolean>, <Numerator>, <Denominator>, <Offset>, <SweepType>
Defines an external generator frequency for frequency-converting measurements. The
external generator frequency is either a range (for frequency sweeps) or a CW fre-
quency (for power, time and CW Mode sweeps).
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the generator is referenced via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
Switch the generator on or off.
<Numerator>
Parameters of the frequency formula. The source frequency f
s
is
calculated according to f
s
= <Numerator>/<Denominator> * f
b
+
<Offset> where f
b
represents the channel base frequency
(parameter
SWEep
). For parameters
CW
or
FIXed
, f
b
= 0.
Note:
The
<Offset>
parameter also includes the "Offset Ratio"
in manual control.
Range:
The converted frequency or frequency range must
be within the external generator's frequency range.
In addition the <Numerator> values are rounded to
positive or negative integer numbers.
*RST:
The default frequency/frequency range corresponds
to the sweep range or CW frequency of the ana-
lyzer, i.e. <Numerator> = <Denominator> = 1, <Off-
set> = 0 Hz.
Default unit: 1
<Denominator>
See above, <Numerator>
Range:
The converted frequency or frequency range must
be within the external generator's frequency range.
In addition the <Denominator> values are rounded
to positive integer numbers; zero is not allowed.
*RST:
1
Default unit: 1
<Offset>
See above, <Numerator>
Range:
The converted frequency or frequency range must
be within the external generator's frequency range.
In addition the <Offset> values are rounded to (pos-
itive or negative) multiples of 1 Hz.
*RST:
0
Default unit: Hz
SCPI Command Reference