
Command Reference
R&S
®
ZNA
1362
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:IFFixed
<FixedFrequency>
Defines a fixed frequency and assigns it to the IF signal (for alternative commands see
example below).
Suffix:
<Ch>
.
Channel number.
Parameters:
<FixedFrequency>
Fixed frequency.
Range:
Depending on the instrument model.
Increment: 0.1 kHz
*RST:
Minimum of the analyzer's frequency range, fMIN
Default unit: Hz
Example:
*RST; FREQ:CONV:MIX:IFF 1 GHz
Reset the analyzer and specify a fixed frequency of 1 GHz, to be
assigned to the IF signal.
FREQ:CONV:MIX:FFIX?; FIX?
Query the fixed frequency and the signal assignment using the
alternative commands. The response is 1000000000;IF.
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOFixed
<FixedFrequency>
Defines a fixed frequency and assigns it to the LO signal (for alternative commands
see example below).
Suffix:
<Ch>
.
Channel number.
Parameters:
<FixedFrequency>
Fixed frequency.
Range:
Depending on the instrument model.
Increment: 0.1 kHz
*RST:
Minimum of the analyzer's frequency range, fMIN
Default unit: Hz
Example:
*RST; FREQ:CONV:MIX:LOF 1 GHz
Reset the analyzer and specify a fixed frequency of 1 GHz, to be
assigned to the LO signal.
FREQ:CONV:MIX:FFIX?; FIX?
Query the fixed frequency and the signal assignment using the
alternative commands. The response is 1000000000;LO.
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:RFFixed
<FixedFrequency>
Defines a fixed frequency and assigns it to the RF signal (for alternative commands
see example below).
Suffix:
<Ch>
.
Channel number.
R&S ZVR/ZVABT Compatible Commands