For mixer measurements, select a "Base Ratio" of 1, a "Frequency Offset" equal to
the LO frequency, and an "Offset Ratio" of ±1. If the RF signal is at the channel
base frequency, the port frequency is at the frequency of the upper (lower) side-
band.
●
To measure n-th order intermodulation products, select integer "Base Ratio" and
"Offset Ratio" with |"Base Ratio"| + |"Offset Ratio"| = n .
●
To measure the n-th harmonic of the channel base frequency, select a "Frequency
Offset" of 0 and an integer "Base Ratio" = n.
Remote command:
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:ARBitrary:IFRequency
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:ARBitrary:EFRequency<Gen>
[SENSe<Ch>:]FREQuency<Pt>:CONVersion:ARBitrary
[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary:PMETer<Pmtr>
Arbitrary Power Tab
Configures the source ports of the R&S
ZNA for port-specific powers.
Channel base power
, all source port powers are set to the channel base power P
b
. For power
sweeps, P
b
corresponds to the unmodified sweep range. For frequency, time, and CW
mode sweeps, it is equal to the fixed "Power". f
b
is defined by the "Stimulus" parame-
ters (see
Chapter 6.8, "Stimulus Softtool"
Active frequency converter ports
are displayed with their RF IN and LO IN compo-
nents.