Command Reference
R&S
®
ZNA
829
User Manual 1178.6462.02 ─ 12
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg>
Segment number
Parameters:
<LimLineType>
LMIN | LMAX | OFF
Limit line type
Range:
LMAX (upper limit line segment), LMIN (lower limit
line segment), OFF (limit check switched off, limit
line segment not deleted)
*RST:
LMAX
Example:
*RST; :CALC:LIM:UPP 0, 0
Define an upper limit line segment across the entire sweep
range, using a constant upper limit of 0 dBm.
CALC:LIM:SEGM:TYPE LMIN
Turn the defined limit line segment into a lower limit line seg-
ment.
Manual operation:
See
CALCulate<Chn>:LIMit:SOUNd[:STATe]
<Boolean>
Switches the acoustic signal (fail beep) on or off. The fail beep is generated each time
the analyzer detects an exceeded limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - Fail beep on or off.
*RST:
OFF
Example:
CALC:LIM:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
Manual operation:
See
CALCulate<Chn>:LIMit:STATe
<Boolean>
Switches the limit check (including upper and lower limits) on or off.
Tip:
Use
CALCulate<Ch>:LIMit:UPPer:STATe
or
CALCulate<Ch>:LIMit:LOWer:STATe
to switch on or off the individual limit checks
for upper or lower limit lines.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - Limit check on or off.
*RST:
OFF
SCPI Command Reference