CALCulate:LIMit:FAIL:ALL?
CALCulate<Chn>:LIMit:STATe:AREA
Limit Fail Beep
Activates or deactivates the fail beep. The fail beep is a low-tone acoustic signal that is
generated each time the analyzer detects an exceeded limit. No fail beep can be gen-
erated if the limit check is switched off.
Note:
The R&S
ZNA does not have a built-in audio device and loudspeaker. To hear
these sounds, connect a USB audio device to the instrument or operate it via remote
desktop.
Remote command:
CALCulate<Chn>:LIMit:SOUNd[:STATe]
Clear Test
Resets the limit check results.
Remote command:
Global Check
Activates or deactivates the global limit check including upper/lower limits and ripple
limits. The global limit check is a composite limit check over all traces of the current
recall set. The result of the global check appears in a popup box whenever "Global
Check" is selected.
●
"PASS" represents pass for all traces with enabled limit check. A trace without limit
lines or with disabled individual limit check always passes the global check.
●
"FAIL" means that the limit check for one or more traces failed.
Remote command:
TTL1 Pass / TTL2 Pass
Assigns the active trace to the low-voltage (3.3 V) TTL output signals at the User Port
(see
Monitoring a single trace
If "TTL1 Pass" ("TTL2 Pass") is selected and the trace passes all active limit checks,
then the TTL signal is applied to pin 13 (pin 14) of the User Port. If one of the limit
checks fails, then no TTL signal is generated.
Monitoring several traces