Concepts and Features
R&S
®
ZNA
237
User Manual 1178.6462.02 ─ 12
To fulfill its purpose, the plug-in can communicate with other processes / peripherals
that are not part of the VNA firmware or host OS.
5.7.30.1
Basic Use Case
A basic use case is the integration of a power supply into an arbitrary VNA sweep.
Suppose that you want to test an amplifier with various bias settings. To do so, you can
perform a CW sweep to measure S-parameters with different gate voltages applied to
the amplifier at each sweep point.
With one power supply channel connected to the amplifier’s gate, and another channel
connected to the amplifier’s drain, the measurement could proceed as follows:
1. At the beginning of the sweep, the power supply is initialized (e.g. voltage setting
and current limit for each power supply channel) and then switched ON.
2. At the beginning of each sweep point, the plugin sets the power supply gate volt-
age to the next value.
3. At the end of each sweep point, the plugin reports the power supply reading of the
gate voltage and drain current to the VNA. These readings are displayed in dedica-
ted traces.
4. At the end of the sweep, the power supply is switched OFF.
5.7.30.2
Technology
The plug-in interface is a traditional C style DLL interface. Suitable DLLs can be com-
piled using different platforms, such as C or MATLAB. A plug-in is installed in the firm-
ware as a
*.zip
file, containing the external DLL and, possibly, additional files that
shall be kept with the DLL.
An SDK for external DLLs is provided with the analyzer firmware on the Rohde&
Schwarz internet pages.
For details, see the HTML help topics that are distributed with the plugin SDK.
Interactions of the external DLL with analyzer firmware and external devices are
synchronized with the measurement. Synchronization is done via callbacks from the
FW, at so-called interrupts. At an interrupt, the measurement takes a break and waits
for the FW to continue. During an interrupt, the plugin can access external devices and
return results to the FW.
The VNA firmware can abort the processing in the external DLL at any time, e.g. on a
[Preset].
The DLL can handle certain trace types that must be uniquely named within the DLL. A
naming scheme for traces allows the firmware (and hence the user) to request a partic-
ular trace.
Optional Extensions and Accessories