GUI Reference
R&S
®
ZNA
342
User Manual 1178.6462.02 ─ 12
The info field shows invalid results ('----') if the wrong sweep type or trace format is
selected, or if no compression point with the configured properties was found.
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe]
CALCulate<Chn>:STATistics:NLINear:COMP:LEVel
CALCulate<Chn>:STATistics:NLINear:COMP:PHASe
CALCulate<Chn>:STATistics:NLINear:COMP:RESult?
CALCulate<Chn>:STATistics[:STATe]:AREA
Reference Value
Selects the reference value ("small signal value") for the compression point calculation.
"First Point"
(default)
uses the value at the first point of the selected evaluation range as
the reference value (see
Chapter 6.5.7.2, "Evaluation Range Dialog"
"Selected
Marker"
uses the value at a selectable marker as the reference value (see
"Range"
uses the average value in a configurable
as the reference value
"Defined
Value"
allows to specify the reference value manually (see
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP:REFerence
Selected Marker
This button is only visible if "Selected Marker" is used as
. It allows
you to select the marker whose value shall be used as the reference ("small signal
value") for the compression point calculation.
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP:RMARker
Ref. Range
This button is only visible if "Range" is used as
. It opens the "Refer-
ence Range" dialog that allows you to configure the sweep range whose average value
shall be used as the reference ("small signal value") for the compression point calcula-
tion.
Traces Softtool