Command Reference
R&S
®
ZNA
889
User Manual 1178.6462.02 ─ 12
For dB formatted traces, the reference level can be set using
STATistics:NLINear:COMP:RLEVel
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Phase>
Reference phase
*RST:
1°
Default unit: deg
Manual operation:
See
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe]
<Boolean>
Displays or hides the compression point result in the diagram area of trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - statistical info field on or off.
*RST:
OFF
Example:
See
CALCulate<Chn>:STATistics:NLINear:COMP:
Manual operation:
See
CALCulate<Chn>:STATistics:RESult?
<Result>
Returns a single statistical parameter of the trace no.
<Chn>
or all parameters. It is not
necessary to display the info field (
CALCulate<Chn>:STATistics[:STATe]
ON
)
before using this command.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Query parameters:
<Result>
MEAN | STDDev | MAX | MIN | RMS | PTPeak | PEAK2p |
ELENgth | PDELay | GAIN | SLOPe | FLATness | ALL
MEAN - return arithmetic mean value of all response values of
the trace in the entire sweep range (or in the evaluation range
defined in manual control).
STDDev - return standard deviation of all response values.
MAX - return the maximum of all response values.
MIN - return the minimum of all response values.
RMS - return the root mean square of all response values.
PTPeak - return the peak-to-peak value (MAX - MIN).
ELENgth - return the electrical length.
PDELay - return the phase delay.
GAIN - return the gain, i.e. the larger of two marker values.
SCPI Command Reference