GUI Reference
R&S
®
ZNA
575
User Manual 1178.6462.02 ─ 12
Table 6-8: Allowed numbers of active ports
"Sym 2x Thru"
other
Single Ended
2
1
Balanced
2 or 4
Tip
: use a 4-port 2x thru when
strong FEXT (far-end crosstalk) is
present
2
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:COUPon[:STATe]
CALCulate:FMODel:SFD<Ph_pt>:COUPon[:STATe]
6.14.5.3
Measure DUT + Test Fixture
The "Measure DUT + Test Fixture" section allows you to measure the whole structure,
i.e. the DUT with the test fixture.
Measure
Starts the measurement of DUT + Test Fixture at the
ports.
The display will change to show all the S parameter measurements being made. The
resulting Touchstone file is written to C:\Users\Public\Documents\Rohde-Schwarz\ZNA
\Embedding.
Remote command:
CALCulate:FMODel:ISD<Ph_pt>:DUT:MEASure
CALCulate:FMODel:SFD<Ph_pt>:DUT:MEASure
Active
, use these checkboxes to indicate the physical ports to
which DUT + test fixture are connected. For single-ended deembedding 2 ports must
be active, for balanced deembedding 4 ports.
Offset Embed Softtool