Command Reference
R&S
®
ZNA
1261
User Manual 1178.6462.02 ─ 12
Example:
SOUR:POW:CORR:TCO:DEL1
Delete the first point in the list.
SOUR:POW:CORR:TCO:DEL:ALL; :SOUR:POW:CORR:TCO?
Clear the entire list. Query whether the transmission coefficients
are still taken into account. The response is
0
.
MMEM:LOAD:CORR:TCO 'C:\Users\Public\Documents
\Rohde-Schwarz\ZNA\Traces\Test.s2p', 'S21'
Load a power loss list from a previously created 2-port Touch-
stone file.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:COUNt?
Queries the number of frequency values and transmission coefficients in the power
loss list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Query only
Manual operation:
See
"Frequency / Transm. Coefficients: Insert, Delete, Delete
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DEFine<ListNo>
<Frequency>, <TransCoeff>
Adds a new frequency and transmission coefficient to the end of the power loss list.
The query returns the frequency and transmission coefficient no.
<ListNo>
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
<ListNo>
Number of point in the list. This suffix is only used for queries.
SCPI Command Reference