GUI Reference
R&S
®
ZNA
345
User Manual 1178.6462.02 ─ 12
●
The analyzer can export the raw complex (unformatted) data or formatted data.
The unformatted data are independent of all "Smooth Shift Hold" settings; see
●
For complex traces, if marker format and trace format do not coincide, the marker
values are calculated before
are applied. In this case, the dis-
played marker values may not behave as expected.
Smoothing
Activates the smoothing function for the active trace, which can be a data or a memory
trace.
With smoothing active, the trace value at each sweep point is replaced by the arith-
metic mean of the trace values at the sweep point itself, and at neighboring sweep
points (symmetrically, to the left and to the right). The number of trace values to be
averaged is defined by the
and can be adjusted according to the properties of
the trace.
Smoothing operates on the formatted trace and does not significantly increase the
measurement time.
Traces Softtool