GUI Reference
R&S
®
ZNA
280
User Manual 1178.6462.02 ─ 12
Figure 6-4: Amplifier Compression dialog
Driving Port/Receiving Port
Selects the driving and receiving port for the gain compression measurement.
Remote command:
[SENSe<Ch>:]FREQuency:COMPression:SRCPort
[SENSe<Ch>:]FREQuency:COMPression:RECeiver
Sweep Type
Compression measurement channels are set up for linear frequency sweeps. For each
point on the (linear) frequency grid, the firmware performs a power sweep.
Start Frequency/Stop Frequency
See the corresponding controls on the
Number of Points
See the corresponding controls on the
Base Power/IF Bandwidth/Step Att Receiving Port
See the corresponding controls on the
.
Start Power/Stop Power
Defines the start and stop power for the power sweeps that are performed at each
point on the linear frequency grid defined using
Start Frequency/Stop Frequency
Remote command:
[SENSe<Ch>:]FREQuency:COMPression:POWer:STARt
[SENSe<Ch>:]FREQuency:COMPression:POWer:STOP
Meas Softtool