GUI Reference
R&S
®
ZNA
266
User Manual 1178.6462.02 ─ 12
Define
Selects the harmonic to be measured.
A "Frequency out of range" warning is displayed the harmonic frequency to be mea-
sured exceeds the frequency range of the analyzer. The harmonic traces are truncated
at the maximum receiver frequency.
"Harmonic"
The Harmonic order can be set in the range between 2 (2nd har-
monic) and 100.
"Receiving
Port"
The receive port measures the signal at
n
times the fundamental fre-
quency where n is the harmonic order.
"Source Port"
The source port provides the stimulus signal at the fundamental fre-
quency.
Two different types of results are provided.
●
In the direct measurement, the nth harmonic of the stimulus signal is measured.
The result is displayed as a wave quantity but can be modified and converted using
the functions in the "Trace" softtool.
●
In the relative measurement, the nth harmonic of the stimulus signal is divided by
1st harmonic (fundamental) received from the DUT. The result corresponds to the
nth harmonic distortion factor.
Meas Softtool