GUI Reference
R&S
®
ZNA
518
User Manual 1178.6462.02 ─ 12
Figure 6-21: Arb Frequency tab, default columns
Active frequency converter ports
are displayed with their RF IN and LO IN compo-
nents.
Non-editable table cells
In addition to the test ports ("Port" or "Conv"), the source ports include all configured
("Source Gen"), and the
port ("Conv. LO"), if not assigned
to a converter.
In addition to the test ports ("Port"), the source ports include all configured external
generators ("Source Gen").
The receive ports include all configured external power meters ("Pmtr" ...).
Each port is displayed with its port number, device type ("Info"), and its frequency
results according to the current channel base frequency and frequency conversion set-
tings.
Values that cannot be modified are displayed with a gray background; frequency
results that cannot be measured are displayed with a red background.
Source RF Off
Turns the RF signal at a source port (analyzer test port or generator port) on or off.
The column is disabled (grayed out) while the RF power is switched off globally for the
active channel. The settings are restored when the RF power is switched on again.
While the RF power is switched off, the port can still be used as a receive port.
Remote command:
SOURce<Ch>:POWer<PhyPt>:STATe
SOURce<Ch>:POWer<PhyPt>:GENerator<Gen>:STATe