GUI Reference
R&S
®
ZNA
457
User Manual 1178.6462.02 ─ 12
The R&S
ZNA always proposes an optimum solution (minimum number of assign-
ments) that also minimizes the physical port reconnections required between calibra-
tion stages. For user-modified assignments, it provides assistive information indicating
insufficient, or redundant entries.
The test port connectors are automatically set according to the connector type of the
selected calibration unit port.
"Detect Port Assignment"
starts a procedure by which the R&S
ZNA (with a little help
from the attached calibration unit) auto-detects the connected ports. The automatic
assignment replaces the configured one.
In case auto-detection fails
●
an error report is shown as a warning dialog
●
the undetected port connections are marked with warning signs
●
the calibration can be invalid
"Default Port Assignment"
restores the default port assignments.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:TPORt
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
"Detect Port Assignment"
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
"Default Port Assignment"
Cal Softtool