[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
<TestPort1>, <CalUnitPort1>, <TestPort2>...
Manually defines port assignment no. <Asg> for channel no. <Ch>.
With manual configuration a non-minimal set of port assignments can be created,
which may increase measurement accuracy. On the other hand, it is up to the user to
ensure that the created set of port assignments is complete and valid for the chosen
calibration type (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Note
that during the corresponding calibration sweep the R&S
ZNA expects the physi-
cal port connections to be established exactly as defined by the port pairs.
Use
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt
command to take advantage of auto-detection of port connections.
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:
to create the default assignments instead.
See
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 156 for background infor-
mation.
Suffix:
<Ch>
.
Number of the channel
<Asg>
Number of the port assignment
Parameters:
<TestPort1>
First test port number
<CalUnitPort1>
Number of the calibration unit port that is assigned to the first
test port
<TestPort2>
Second test port number ...