Command Reference
R&S
®
ZNA
1097
User Manual 1178.6462.02 ─ 12
CPUosm
PFNPort (PUOSM) Scalar Vector Mixer calibration
<Characterization>
Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
–
The empty string (
' '
) refers to the factory calibration of (and
stored on) the active calibration unit.
–
A VNA cal kit file name
*.calkit
without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
Example:
SENSe1:CORRection:COLLect:AUTO:CONFigure
FNPort,''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
Example:
See
Chapter 9.2.5.5, "Vector Mixer Calibration Using Calibration
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
<PortNumber>
Defines the "Common Port" of an automatic calibration.
This port will be part of every port assignment that is defined using
]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
If
[SENSe:]CORRection:COLLect:AUTO:MCONnect
ON
, this port will be used as
the center of all Through measurements. If you set up the port assignments manually,
make sure that every port assignment contains it.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<PortNumber>
Number of the common port
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:MIXer
<Port1>, <Port2>
Acquire calibration data for the calibration mixer in an automatic vector mixer calibra-
tion
SCPI Command Reference