"Delta F"
sweep whose standard parame-
ters can be set in the dialog (see
"CW Frequency/Number of Points/
The frequency distance between lower tone and upper tone starts at
"Start Delta Freq." and ends at "Stop Delta Freq." (grows or shrinks
linearly). Either the lower tone or the upper tone remains at the con-
figured CW Frequency ("CW fixed to": "Lower Tone"/"Upper Tone").
As a third possibility, the center between lower tone and upper tone
can remain at the configured CW Frequency ("CW fixed to": "Center".
For the related parameters, see
"Start Delta Frequency/Stop Delta
Start Frequency/Stop Frequency/Number of Points/Base Power
Standard parameters of a linear frequency sweep (see
Start Power/Stop Power/Number of Points/Fixed CW
Standard parameters of a power sweep (see
Chapter 6.8, "Stimulus Softtool"
Start Delta Frequency/Stop Delta Frequency/CW fixed to
Defines the lower tone and upper tone frequency sweeps for
F".
Delta Frequency
Defines the frequency difference between the upper and the lower tone. The tone dis-
tance is kept constant across the entire sweep.
Remote command:
[SENSe<Ch>:]FREQuency:IMODulation:TDIStance
CW Frequency/Number of Points/Base Power
Standard parameters of a "CW Mode" sweep (see
Chapter 6.8, "Stimulus Softtool"
Additional Channel Settings
IF Bandwidth
See
Step Att Receiving Port
Only available if the selected
is equipped with a receiver step attenua-
Chapter 5.7.21, "Receiver Step Attenuators"
at this port.
Frequency Overview Dialog
The "Frequency Overview" dialog summarizes the resulting measured frequency
ranges, according to the current setup of the related channel. Invalid frequency ranges
(that are not covered by the analyzer's frequency range) are marked in red.
Meas Softtool