Getting Started
R&S
®
ZNA
12
User Manual 1178.6462.02 ─ 12
Risk of injury if feet are folded out
The feet can fold in if they are not folded out completely or if the instrument is shifted.
Collapsing feet can cause injury or damage the instrument.
●
Fold the feet completely in or out to ensure stability of the instrument. Never shift
the instrument when the feet are folded out.
●
When the feet are folded out, do not work under the instrument or place anything
underneath.
●
The feet can break if they are overloaded. The overall load on the folded-out feet
must not exceed 500
N.
4.1.4
Operation in a 19" Rack
The R&S
ZNA can be mounted in 19" racks using the adapter R&S
ZZA-KN6 (order
number 1332.4498.02). Proceed according to the mounting instructions supplied with
the rack adapter.
Risk of instrument damage due to insufficient airflow in a rack
If you mount several instruments in a rack, you need an efficient ventilation concept to
ensure that the instruments do not overheat. Insufficient airflow for a longer period can
disturb the operation and even cause damage.
4.1.5
EMI Suppression
Electromagnetic Interference (EMI) can affect the measurement results.
To suppress generated Electromagnetic Interference:
●
Use suitable shielded cables of high quality (see table below)
●
Always terminate open cable ends
●
Note the EMC classification in the data sheet
Putting the Analyzer into Operation