GUI Reference
R&S
®
ZNA
263
User Manual 1178.6462.02 ─ 12
The notation for ratios follows the usual scheme of the vector network analyzer:
●
The a-waves are the outgoing/transmitted waves at the analyzer's test ports.
●
The b-waves are the incoming/measured waves.
●
The source port for the stimulus signal must be specified in addition.
●
The port number range covers all test ports of the analyzer.
Numerator
Selects the type (left pull-down list) and the port number assignment (right pull-down
list) of the wave that forms the numerator of the
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "B2/A1" | ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "B2/A1" | ...
Denominator
Selects the type (left pull-down list) and the port number assignment (right pull-down
list) of the wave that forms the denominator of the
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "B2/A1" | ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "B2/A1" | ...
Use Generator as Source
If selected, the "Port" control lists the external generators. Otherwise it lists the ana-
lyzer ports.
External generators must be configured explicitly in the "External Generators" dialog
before they appear in the list. See also
Chapter 5.7.28, "External Generators"
Meas Softtool