Figure 6-19: Define Calibration dialog: automatic calibration
The green arrow symbols give a preview of the type and the number of calibration
sweeps involved:
●
Curved arrows (example: "Refl Norm Open") denote one or more reflection mea-
surements at each port.
Reflection calibration types can be used for any set of test ports: reflection calibra-
tions are repeated for each port.
●
Straight, horizontal arrows (example: "Trans Norm") denote one or more transmis-
sion measurements between each pair of two ports.
Transmission calibration types require at least two physical ports. For unidirectional
transmission calibration types ("Trans Norm", "One Path Two Ports"), the
(
"Source"
port) must be specified in addition.
●
The full n-port calibration types (n > 1, e.g. "TOSM") are symbolized by a closed
square symbol. The number of arrows increases the complexity but can also
improve the accuracy of the calibration.
For an overview, refer to
Remote command:
Manual calibration
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
Automatic calibration
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
CalUnit Con ← Cal. Settings
Opens the "Calibration Unit" dialog that allows you to define the port assignments for
an automatic calibration.
Cal Softtool