[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalName>, <CalType>,
<TestPort1>[, <TestPort2>, <TestPort3orAdapterCalKit>, <TestPorts>...]
Defines the calibration to be performed for channel
<Ch>
.
If multiple channels shall be calibrated in parallel, use
●
[SENSe:]CORRection:COLLect:CHANnels:ALL
●
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
before executing this command. For background information see
lel Calibration of Multiple Channels"
Calibration data is acquired using
[SENSe<Ch>:]CORRection:COLLect[:
.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName>
Name of the calibration (string parameter). The name serves as
a reference to delete a particular set of system correction data
(
[SENSe<Ch>:]CORRection:COLLect:DELete
<CalType>
ARTosm | ARPTosm | REFL | RSHort | FOPort | FTRans |
RTRans | FRTRans | OPTPort | TOSM | UOSM | VUOSm |
TRL | TOM | TSM | TRM | TNA | PTOSm | PTOM | PTSM |
PTRM | PTRL | PTNA | PUOSm | VPUosm | CPUosm |
MTOSm | MUOSm | MOSM | MTRL
Calibration types; see list below.
<TestPort1>
First calibrated test port number. For an n-port calibration type, n
port numbers must be specified. If more than n numbers are
defined, the spare numbers (the last ones in the list) are ignored.
Entering less than n numbers causes an error message. For a
one path two port calibration (
OPTPort
), the first port no.
denotes the node port (fully calibrated port).
<TestPort2>
Second calibrated port number.
<TestPort3 or
AdapterCalKit>
Either the third calibrated port number, or - for Adapter Removal
calibrations only - the file path (relative to the current directory)
where the analyzer shall store the adapter characterization
obtained during the calibration.
The latter is optional. If a file path is provided the characteriza-
tion file (in Touchstone s2p format) is generated when the cor-
rection terms are finally calculated (
]CORRection:COLLect:SAVE:SELected[:DUMMy]
<TestPorts>
More ports to be calibrated.