Command Reference
R&S
®
ZNA
1089
User Manual 1178.6462.02 ─ 12
Manual operation:
See
[SENSe:]CORRection:COLLect:AVERage
<Average>
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to manual aver-
aging (see
) the number of calibration sweeps is
calculated automatically.
Parameters:
<Average>
AUTO | MANual
Manual operation:
See
[SENSe:]CORRection:COLLect:CHANnels:ALL
<Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
●
Manual SMARterCal
●
Automatic SMARTerCal (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean>
ON
–
calibrate all channels.
OFF
–
calibrate the active channel only.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
<Boolean>
Toggles the
M
ultiple
C
alibration
TYpes
mode that allows you to calibrate a subset of
the available channels using channel-specific ports and calibration types (see
ter 5.5.8, "Parallel Calibration of Multiple Channels"
Enable the MCTYpes mode before defining the calibrations to be performed using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
. Disable the MCTYpes
mode after the calibrations were saved using
. The latter must be called only
once
, with <Ch>
referring to one of the channels to be calibrated (see the example below).
Parameters:
<Boolean>
SCPI Command Reference