GUI Reference
R&S
®
ZNA
451
User Manual 1178.6462.02 ─ 12
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
Detect Ports & Start Cal
Performs the "Detect Port Assignment" and "Start Cal Sweep" functions, one after the
other.
Detect Port Assignment
Starts a procedure by which the R&S
ZNA (with a little help from the attached calibra-
tion unit) auto-detects the connected ports. The automatic assignment replaces the
configured one.
In case auto-detection fails
●
an error report is shown as a warning dialog
●
the undetected port connections are marked with warning signs
●
the calibration can be invalid
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Apply
This button is enabled when sufficient data have been acquired for the calibrated ports.
The button starts the calculation of the calibration data and closes the calibration wiz-
ard.
The results are applied to the related channels. The current instrument settings are
stored with the correction data.
To avoid incompatibilities, older calibration data is deleted unless it has been transfer-
red into the calibration "Pool" using the
Note: Checks during the calculation of correction data
Incompatibilities between the selected calibration type, the standards and the channel
settings can cause the calibration to be inaccurate. The analyzer auto-detects potential
sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
Cal Softtool