Command Reference
R&S
®
ZNA
1098
User Manual 1178.6462.02 ─ 12
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Port1>
VNA port connected to the RF port of the calibration mixer
<Port2>
VNA port connected to the IF port of the calibration mixer
Example:
See
Chapter 9.2.5.5, "Vector Mixer Calibration Using Calibration
Usage:
Setting only
Options:
R&S
ZNA-K5
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
<Characterization>,
<TestPort1>, <CalUnitPort1>, <TestPort2>...
Selects and initiates an automatic calibration at arbitrary analyzer and calibration unit
ports. A progress monitor for the calibration sweeps is displayed.
Tip:
●
This command is necessary if the analyzer fails to detect the cal unit ports connec-
ted to each of its ports (e.g. because of a high attenuation in the test setup). If
auto-detection works you can use the simpler command
●
If several calibration units are connected, use
to select a unit for the calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<Characterization>
Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
–
If an empty string (
' '
) is specified, the factory cal kit file
stored in the active calibration unit is used.
–
A VNA cal kit file name
*.calkit
without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Test port numbers. For an n-port automatic calibration, n arbi-
trary (not necessarily consecutive) port numbers must be speci-
fied.
SCPI Command Reference