GUI Reference
R&S
®
ZNA
566
User Manual 1178.6462.02 ─ 12
Table 6-6: Direct Compensation data
Prompt for Each Port
File Type
Description
disabled
csv
One csv trace file per standard,
containing reflection traces for all
selected ports; see
ter 5.4.2.2, "ASCII (*.csv) Files"
enabled
s1p
One 1-port Touchstone file per
standard and port (see
ter 5.4.2.1, "Touchstone Files"
Tip:
Remote control provides additional flexibility. You can:
●
Measure the same ports repeatedly without changing the standards and attribute
the results to different channels.
●
Calculate the compensation data for different ports, using mixed Open and Short
standards.
Refer to the program example for
[SENSe<Ch>:]CORRection:COLLect:
.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
[SENSe<Ch>:]CORRection:COLLect:FIXTure:SAVE
[SENSe<Ch>:]CORRection:COLLect:FIXTure:EXPort
[SENSe<Ch>:]CORRection:COLLect:FIXTure:IMPort
6.14.3
One Way Loss Tab
Defines loss parameters for each physical port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see
Chapter 6.14.1.2, "Offset Panel"
Background information
Refer to the following sections.
●
Chapter 5.6, "Offset Parameters and De-/Embedding"
●
Chapter 5.6.1.2, "Definition of Loss Parameters "
●
Chapter 5.6.1.4, "Auto Length and Loss"
●
Chapter 5.6.1.5, "Fixture Compensation "
Offset Embed Softtool