Command Reference
R&S
®
ZNA
1107
User Manual 1178.6462.02 ─ 12
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalName>
Name of the calibration (string parameter) defined together with
the calibration type (
[SENSe<Ch>:]CORRection:COLLect:
ALL
- the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of
*RST:
-
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
CORR:COLL:DEL ALL
Calculate the system error correction data and apply it to the
active channel, then delete the data.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:DISCard
Terminates a system error correction, discarding the acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
:SENSe1:CORRection:COLLect:METHod:DEFine 'Test', TOSM, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected THRough, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected SHORt, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 2
:SENSe1:CORRection:COLLect:DISCard
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
<StandardType>,
<TestPort1>, <TestPort2>...
Starts a fixture compensation sweep in order to acquire measurement data for a test
fixture that has its inner conductor terminated with the selected standards.
SCPI Command Reference