GUI Reference
R&S
®
ZNA
480
User Manual 1178.6462.02 ─ 12
ibration kit file, instead of the gender. During the calibration, the analyzer checks
whether the cal kit contains the necessary standard types for the required ports.
Remote command:
[SENSe:]CORRection:CKIT:<StandardType>
[SENSe:]CORRection:CKIT:<ConnType>:SELect
Add / Copy... / Delete / View / Modify...
The buttons in the right part of the dialog are used to manage standards:
●
"Add" adds a new standard to the calibration kit. The properties of the standard can
be edited in the table.
●
"Copy..." creates a standard based on the properties of an existing standard.
●
"Delete" deletes the selected standard.
●
"View / Modify..." opens the "View / Modify Cal Kit Standards" dialog. This dialog
shows the circuit model for the selected standard. For user-defined standard, you
can modify the circuit model parameters. See
"View / Modify Cal Kit Standards Dia-
Remote command:
[SENSe:]CORRection:CKIT:<StandardType>
[SENSe:]CORRection:CKIT:<StandardType>:WLABel
Read .s1p File... / Read .s2p File...
Opens a file selection dialog where you can select a Touchstone file containing the
reflection or transmission S-parameters for the standard. The R&S
ZNA uses the
imported S-parameters rather than the circuit model to characterize the standard, if
".s1p File"/".s2p File" is checked in the standard table. The appropriate file type (
*.s1p
for one-port standards and
*.s2p
for two-port standards) is selected automatically.
Remote command:
View / Modify Cal Kit Standards Dialog
The "View / Modify Cal Kit Standards" dialog shows the circuit model for a selected cal-
ibration standard. It is also used to define or edit the circuit model (offset and load)
parameters for a user-defined standard.
Related information
Refer to the following sections:
●
Chapter 5.5.2.4, "Cal Kit Files"
●
●
Chapter 5.5.2.1, "Calibration Standard Types"
Access:
> "View / Modify..."
Cal Softtool