Command Reference
R&S
®
ZNA
1140
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]FREQuency[:CW]
<FixedFreq>
[SENSe<Ch>:]FREQuency:FIXed
<FixedFreq>
Defines the fixed (Continuous Wave, CW) frequency for all sweep types operating at
fixed frequency ("Power", "Time", "CW Mode"). The two command forms
[SENSe<Ch>:]FREQuency[:CW]
and
[SENSe<Ch>:]FREQuency:FIXed
are
equivalent.
This command also defines the center frequency of the intermodulation spectrum dia-
gram (with option R&S
ZNA-K4). The frequency range depends on the R&S
ZNA
model; see
Chapter 8.3.14.12, "[SENSe:]FREQuency..."
Note:
[SENSe<Ch>:]FREQuency[:CW]|:FIXed
is equivalent to
SOURce<Ch>:FREQuency<PhyPt>[:CW]|:FIXed
. Source and receiver frequency
are always equal; the four commands overwrite each other.
With option R&S
ZNA-K4, Frequency Conversion Measurements, port-specific source
and receiver frequencies can be defined; see
CONVersion:ARBitrary:IFRequency
.
Suffix:
<Ch>
.
Channel number
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency
*RST:
1 GHz
Default unit: Hz
Example:
FUNC "XTIMe:POW:A1"
Activate a time sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation:
See
[SENSe<Ch>:]FREQuency:COMPression:POINt
<CompVal>
Defines the compression value for a gain compression measurement in channel <Ch>.
The gain compression measurement searches for the <CompVal>
dB compression
points.
Suffix:
<Ch>
.
Channel number
Parameters:
<CompVal>
Compression value
SCPI Command Reference