[SENSe<Ch>:]SWEep:DWELl:IPOint
<InsertionPoints>
Defines whether the measurement delay (previously defined via
) is inserted before all partial measurements or before the first partial
measurement only.
Suffix:
<Ch>
.
Channel number
Parameters:
<InsertionPoints>
ALL | FIRSt
Insertion before all or before the first partial measurement
*RST:
ALL
Example:
See
Manual operation:
See
"All Partial Meas'ments / First Partial Meas'ment"
[SENSe<Ch>:]SWEep:POINts
<SweepPoint>
Defines the total number of measurement points per sweep ("Number of Points").
Values between 1 and 100,001 can be set.
Suffix:
<Ch>
.
Channel number 1
Parameters:
<SweepPoint>
Number of points per sweep
Range:
see above
*RST:
201
Example:
*RST
Reset the instrument, activating a frequency sweep with 201
sweep points.
SWE:TIME?
Query total sweep time.
SWE:POIN 2010
Multiply the (default) number of points by 10.
SWE:TIME?
Query total sweep time again. The analyzer estimates a sweep
time that is also multiplied by 10.
Manual operation:
See
[SENSe<Ch>:]SWEep:REVerse
<Boolean>
Reverses the direction of the sweep in channel <Ch>.
Suffix:
<Ch>
.
Channel number