Command Reference
R&S
®
ZNA
1211
User Manual 1178.6462.02 ─ 12
Parameters:
<Boolean>
*RST:
OFF
Manual operation:
See
[SENSe<Ch>:]SWEep:STEP
<StepSize>
Sets the distance between two consecutive sweep points in a (non-segmented) linear
frequency sweep.
Suffix:
<Ch>
.
Channel number
Parameters:
<StepSize>
Frequency step size.
The step size is equal to the current sweep spandivided by the
number of sweep points minus one (see
,
respectively). See also the description of manual control and the
program example below.
Range:
Depends on the current sweep span and the maxi-
mum number of sweep points: "Span Frequency" /
(max. "Number of Points" – 1) ≤ "Freq Step Size" ≤
"Span Frequency"
*RST:
Depends on the analyzer model. The default step
size is equal to the default sweep span of the ana-
lyzer divided by the default number of sweep points
minus one.
Default unit: Hz
Example:
*RST; :SWE:STEP?
Query the default step size. Currently for all analyzers the
default sweep span is the full frequency range and the default
number of points is 201. Hence the response is (max. frequency
- min. frequency) / 200.
SWE:STEP UP
Increase the step size.
FREQ:STOP?; :SWE:POIN?
Query the stop frequency of the sweep and the number of
points. Increasing the step size has changed both values.
Manual operation:
See
[SENSe<Ch>:]SWEep:TIME
<SweepDuration>
Sets the duration of the sweep ("Sweep Time"). Setting a sweep time disables the
automatic calculation of the (minimum) sweep time; see
.
Note:
The sweep time is ignored for the sweep types "Time" and "CW Mode"
).
SCPI Command Reference