Command Reference
R&S
®
ZNA
899
User Manual 1178.6462.02 ─ 12
Manual operation:
See
"Time Start / Time Stop / Time Center / Time Span"
Note:
If the x-axis is scaled in distance units (
CALCulate<Chn>:TRANsform:TIME:
DISTance
), then the stop value is entered in m; the range and default value
changes accordingly.
CALCulate<Chn>:TRANsform:TIME:WINDow
<WindowType>
Selects the window type for filtering the data in the frequency domain prior to the time
domain transformation.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<WindowType>
RECT | HAMMing | HANNing | BOHMan | DCHebyshev
RECT - no profiling (rectangle)
HANN - normal profile (Hann)
HAMMing - low first sidelobe (Hamming)
BOHMan - steep falloff (Bohman)
DCHebyshev - arbitrary sidelobes (Dolph-Chebychev)
*RST:
HANN
Example:
See
CALCulate<Chn>:TRANsform:TIME:DCHebyshev
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:XAXis
<Unit>
Switches over between the x-axis scaling in time units or distance units.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Unit>
TIME | DISTance
TIME - x-axis scaled in time units.
DISTance - x-axis scaled in distance units (Distance = Time * c
0
* Velocity Factor).
Example:
*RST; :CALC:TRAN:TIME:STAT ON
Reset the instrument, activating a frequency sweep, and enable
the time domain transformation for the default trace.
CALC:TRAN:TIME:XAX DIST
Convert the x-axis scaling to distance units.
Manual operation:
See
SCPI Command Reference