Programming Examples
R&S
®
ZNA
1402
User Manual 1178.6462.02 ─ 12
// VPUOSM vector mixer calibration with calu's factory characterization
:SENSe1:CORRection:COLLect:AUTO:CONFigure VPUosm, ''
// connect the RF port (port 1) to calu port 1
// connect the IF port (port 2) to calu port 2
// declare the port assignment
:SENSe1:CORRECTION:COLLECT:AUTO:ASSignment1:DEFine 1,1, 2,2
// Note: the following steps can be performed in any order
// acquire UOSM calibration data between RF and IF port
// currently, this must be the first calibration step
:SENSE1:CORRECTION:COLLECT:AUTO:ASSignment1:ACQUire
// connect the power meter to the RF port (port 1)
// and perform a receiver power calibration
:SENSe1:CORRection:COLLect:AUTO:POWer 1
// connect the calibration mixer and acquire calibration data
:SENSe1:CORRection:COLLect:AUTO:MIXer 1,2
This also works, but mind the order!
:SENSe1:CORRection:COLLect:AUTO:PORTs:TYPE VPUosm, '', 1,1, 2,2
:SENSe1:CORRection:COLLect:AUTO:MIXer 1,2
// receiver power cal must be last step!
:SENSe1:CORRection:COLLect:AUTO:POWer 1
CPUOSM Calibration
A SMARTer UOSM scalar vector mixer calibration using a calibration unit, a power
meter and a calibration mixer.
Same procedure as described in
VPUosm
by
CPUosm
.
9.2.6
Mixer Measurement
The following example is a remote implementation of the tasks the
The power meter and a four-port calibration unit has to be configured beforehand.
// Reset the analyzer
*RST
// Define the mixer
SENSe1:SWEEp:TYPE LINear
SENSe1:SWEEp:POINts 201
SENSe1:FREQuency1:STARt 500 e6 // start : 500 MHz
SENSe1:FREQuency1:STOP 2.5 e9 // stop : 2.5 GHz
SOURce1:POWer1 0 // Base power
SENSe1:FREQuency1:CONVersion:MIXer:FIXed LO
SENSe1:FREQuency1:CONVersion:MIXer:FUND RF
Condensed Programming Examples