GUI Reference
R&S
®
ZNA
588
User Manual 1178.6462.02 ─ 12
Use the complementary dock widget to activate or deactivate ground loop dembedding
and embedding (see
Chapter 6.14.1.6, "Ground Loop Panel"
If
are configured and
enabled, use the dock widget to select the port group whose ground loop network shall
be configured.
Background information
Refer to
Chapter 5.6.2.7, "Ground Loop De-/Embedding"
If the "Fixture Simulator" is disabled for the related channel (see
on page 514), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
See
Type
Switches between "Deembedding" and "Embedding" network definition.
Network
The graphical list contains all available 1-port networks:
The symbol selects "no network" and disables ground loop de-/embedding.
The "1-Port Data" network is defined by imported S-parameter data; see
The remaining networks (Shunt L and Shunt C) are defined by lumped elements
whose parameters are displayed below the graphical list.
Tip:
●
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol.
Offset Embed Softtool