GUI Reference
R&S
®
ZNA
557
User Manual 1178.6462.02 ─ 12
This panel can also be activated by selecting the
for background information, parameters and additional remote commands.
Active
The checkboxes in the "Active" column activate or deactivate the configured de-/
embeding for the related balanced port (i.e. adds or removes it to/from the calculation
flow) without changing its parameters.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<LogPt>[:
STATe]
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>[:
STATe]
File Name <i>/Swap Gates
The ellipsis button in the "File Name <i>" column is enabled as long as the selected
comprises a two-port or four-port data network (
*.s2p
or
*.s4p
file).
When loading a touchstone file, the analyzer by default assumes odd ports left (VNA
side), even ports right (DUT side). However, it is also possible to instruct the analyzer
to assume an "Increasing Port Sequence", i.e. that the low port numbers are on the
VNA side. Furthermore it is possible to "Swap Gates", i.e. to instruct the analyzer to
swap the VNA and DUT side.
Note:
The loaded S-parameters are stored in the active recall set. Persisted recall sets
contain the full embedding and deembedding data so that they can be transferred to
other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:BALanced:DEEMbedding<LogPt>
MMEMory:LOAD:VNETworks<Ch>:BALanced:EMBedding<LogPt>
Balanced Ports
that allows you to modify the balanced
port configuration.
6.14.1.6
Ground Loop Panel
Allows you to activate or deactivate ground loop deembedding/embedding ("Active").
Offset Embed Softtool