Command Reference
R&S
®
ZNA
1147
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:FIXed<Stg>
<InputOutput>
Selects the mixer input or output signal which is at the fixed frequency defined via
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:MFFixed
.
Suffix:
<Ch>
.
Channel number.
<Stg>
Counter for fixed signal (1 if only one mixer stage is used, 1 or 2
for 2 mixer stages).
Parameters:
<InputOutput>
RF | LO | LO1 | LO2 | IF
RF - mixer input signal
LO | LO1 | LO2 - local oscillator signal 1 or 2 (for 2-stage mixer
measurements, see
IF - mixer output signal (mixing product)
*RST:
LO
Example:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:FUNDamental
<InputOutput>
Selects the mixer input or output signal which is at the fundamental frequency (channel
frequency, to be defined by means of
[SENSe<Ch>:]FREQuency:STARt
,
[SENSe<Ch>:]FREQuency:STOP
,
[SENSe<Ch>:]FREQuency:CW
etc.).
Suffix:
<Ch>
.
Channel number.
Parameters:
<InputOutput>
RF | LO | LO1 | LO2 | IF
RF - mixer input signal
LO | LO1 | LO2 - local oscillator signal 1 or 2 (for 2-stage mixer
measurements; see
IF - mixer output signal (mixing product)
*RST:
RF
Example:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
Manual operation:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:IFPort
<PortNumber>
Selects an analyzer port as receive port for the IF signal.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference