GUI Reference
R&S
®
ZNA
565
User Manual 1178.6462.02 ─ 12
Figure 6-28: Measure Fixture dialog: Auto Length (and Loss)
To acquire the necessary data, proceed as indicated in the information area.
With Prompt for Each Port disabled, "Take" acquires data for all selected ports in one
go. Otherwise sweeps are taken port by port.
For "Direct Compensation", it is also possible to save the acquired data to file. In future
measurements, you can load these files instead of repeating the data acquisition.
Figure 6-29: Measure Fixture dialog: Direct Compensation (and Prompt for Each Port)
"Direct Compensation" data files are standard trace files, containing reflection parame-
ter traces for the related port and standard:
Offset Embed Softtool