Command Reference
R&S
®
ZNA
943
User Manual 1178.6462.02 ─ 12
<Cmp>
Number of resistance in circuit model. The total number of resis-
tances depends on the selected circuit model.
Parameters:
<Resistance>
Resistance R<Cmp> for the specified circuit model.
Range:
-10 MΩ to 10 MΩ.
Increment: 1 mΩ (1E-3 Ω)
*RST:
0 Ω for all resistances connected in series with an
inductance. 10 MΩ for all resistances connected in
parallel with a capacitance
Default unit: Ohm
Parameters for setting and query:
<CircuitModel>
CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL | SHLC | GSL |
LSG | SLG | SGL
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:R1? CSL;
R2? CSL
Query the default resistances for the Serial C, shunt L circuit
model. The response is 10000000; 0.
CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3
Increase the resistance R2 to 2.2 kΩ.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition
<CircuitModel>
Selects the circuit model for single ended port deembedding.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<CircuitModel>
FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL |
SHLC | GSL | LSG | GSG | SLG | SGL | SGG
Possible circuit models (character data); see
*RST:
CSL
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>[:STATe]
<Boolean>
Enables or disables the embedding function for single ended ports. It is allowed to
change the circuit model and its parameters while embedding is enabled.
SCPI Command Reference