Command Reference
R&S
®
ZNA
939
User Manual 1178.6462.02 ─ 12
Use
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding:DELete
/
CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<ListId>:DELete
on page 935 to delete all port sets (including port pairs).
Suffix:
<Ch>
.
Channel number
<ListId>
Index of the defined port set within the channel's overall list of
port sets for deembedding/embedding.
Parameters:
<Port1>, <Port2>, ... A port set consist of two or more (different) ports.
A port can be an element of multiple port sets.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>[:
STATe]
<Boolean>
Enables or disables the deembedding function for single ended ports. It is allowed to
change the circuit model and its parameters while deembedding is enabled.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<Boolean>
ON - deembedding active
OFF - deembedding inactive
*RST:
OFF
Example:
CALC:TRAN:VNET:SEND:DEEM:TND CSL
Select the Serial C, shunt L circuit model for deembedding.
CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3; :
CALC:TRAN:VNET:SEND:DEEM ON
Increase the resistance R2 for the Serial C, shunt L circuit model
to 2.2 kΩ and enable deembedding.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
PARameters:DATA
<Interchange>, <arg1>
Defines an embedding network for a single-ended port based on the given S-Parame-
ter traces.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<PhyPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
cuit data from a Touchstone file located at the R&S
ZNA's file system.
SCPI Command Reference