Concepts and Features
R&S
®
ZNA
152
User Manual 1178.6462.02 ─ 12
Advantages of automatic calibration
Automatic calibration is faster and more secure than manual calibration, because:
●
There is no need to connect several standards manually. The number of connec-
tions to be performed quickly increases with the number of ports.
●
Invalid calibrations due to operator errors (e.g. wrong standards or improper con-
nections) are almost excluded.
●
No need to handle calibration kit data.
●
The internal standards do not wear out because they are switched electronically.
Limitations of automatic calibration
Calibration units do not contain characterized Through standards, so calibration types
TOM, TSM, TRM, TRL, TNA are not available.
Safety instructions
Please observe the safety instructions provided with the calibration unit to avoid any
damage to the unit and the network analyzer. Safety-related aspects of the connection
and operation of the units are also reported in the following sections.
Calibration