Command Reference
R&S
®
ZNA
1109
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]CORRection:COLLect:FIXTure:EXPort
<FixtureFile>,
<StandardType>, <TestPort1>, <TestPort2>...
[SENSe<Ch>:]CORRection:COLLect:FIXTure:IMPort
<FixtureFile>,
<StandardType>, <TestPort1>[, <TestPort2>...]
Loads/saves "Direct Compensation" data from/to the specified file.
The
EXPort
command first acquires the required data, just as a
]CORRection:COLLect:FIXTure[:ACQuire]
would do.
"Direct Compensation" data files are standard
, containing reflection param-
eter traces for the related port(s) and standard.
Table 8-12: Direct Compensation data
#TestPorts
File Type
1
s1p
>=1
csv
For 1-port Touchstone files (*.s1p) only a single test port can be specified - otherwise
an error is raised.
Suffix:
<Ch>
.
Channel number
Parameters:
<FixtureFile>
Path to the "Direct Compensation" data file, either absolute or
relative to the current directory (set/queried using
Setting parameters:
<StandardType>
OPEN | SHORt
Terminating standard type
<TestPort1>[,
<TestPort2>, ...]
Test port numbers. For a fixture compensation, n different (but
not necessarily consecutive) port numbers must be specified.
Example:
*RST; SENSe1:CORR:COLL:FIXT:LMP OFF
Activate "Direct Compensation".
SENSe1:CORR:COLL:FIXT:EXP 'Traces\p12_short.csv', SHOR, 1, 2
Acquire "Direct Compensation" data for ports 1 and 2 (whose
inner connectors have to be terminated with a Short) and save
the acquired data to file.
In a subsequent measurement session you can load and apply
these data as follows:
*RST;
:SENSe1:CORR:COLL:FIXT:LMP OFF
:SENSe1:CORR:COLL:FIXT:IMP 'Traces\p12short.csv', SHOR, 1, 2
:SENSe1:CORR:COLL:FIXT:SAVE
Usage:
Setting only