HI 99151

pH/temperature Meter for the Beer Industry

Quick Reference Guide

MEASUREMENTS

Range

:

-2.00 to 16.00 pH
-5.0 to 105.0 °C / 23.0 to 221.0 °F

Accuracy

:

±0.02 pH

(@25 °C/77 °F)

±0.5 °C up to 60 °C; ±1.0 °C outside
±1.0 °F up to 140 °F; ±2.0 °F outside

OPERATIONAL GUIDE

To switch ON and OFF

: press 

ON/OFF/MODE

.

To freeze the display:

 press 

SET/HOLD

 until “

HOLD

“ appears.

To take measurements: 

connect the probe when 

the meter is off

,

remove the protective cap, turn the meter ON, insert the probe tip in the
sample and wait until the “

Not Stable

” tag is turned off.

To enter the pH Calibration mode

: press 

ON/OFF/MODE

 until “

CAL

appears. Release the button and “

pH 7.01 USE

” (or “

pH 6.86 USE

”)

message is displayed.

To select °C or °F unit

: press 

ON/OFF/MODE

 until “

TEMP

“ appears,

then select the temperature unit with 

SET/HOLD

. Press 

ON/OFF/MODE

twice to return to normal mode.

To change the calibration buffer set:

 press 

ON/OFF/MODE

 until

TEMP

“ appears. Press again 

ON/OFF/MODE

: the message “

pH 7.01

BUFF

” (or “

pH 6.86 BUFF

”) appears. Change the buffer set with 

SET/

HOLD

. Press 

ON/OFF/MODE

 to return to normal mode.

w w w . h a n n a i n s t . c o m