Command Reference
R&S
®
ZNA
1361
User Manual 1178.6462.02 ─ 12
Calibration type
Parameter in
[SENSe<Ch>:]CORRec-
tion:COLLect:METHod
Available error terms (depending on port
numbers)
One-port normalization
(reflection) using an open
standard
REFL1
REFL2
REFL12
'SCORR3'
'SCORR9'
'SCORR3' and 'SCORR9'
Full one port
FOPort1
FOPort2
FOPort12
'SCORR1' to 'SCORR3'
'SCORR7' to 'SCORR9'
'SCORR1' to 'SCORR3' and 'SCORR7' to
'SCORR9'
Two-port normalization
FTRans
RTRans
FRTRans
'SCORR6'
'SCORR12'
'SCORR6' and 'SCORR12'
One path two port
FOPTport
ROPTport
'SCORR1' to 'SCORR3', 'SCORR6'
'SCORR7' to 'SCORR9', 'SCORR12'
TOSM
TOSM
'SCORR1' to 'SCORR12'
(at present the isolation terms 'SCORR4' and
'SCORR10' are not included)
TOM, TSM, TRM, TRL,
TNA
TOM | TRM | TRL | TNA
'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK' (for reading
and writing)
'G11' ... 'G22' and 'H11', 'H12', 'H22' (for read-
ing only; the 'H21' matrix elements are nor-
malized to 1)
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:MAGNitude
<OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Tip:
Use the
[SENSe<Ch>:]CORRection:LOSS<PhyPt>...
commands to define
the complete set of loss offset parameters.
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:MAGNitude
is equivalent to
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet
.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<OffsetLoss>
Frequency-independent part of the offset loss
Range:
-200 dB to +200 dB
Increment: 0.0001 dB
*RST:
0 dB
Default unit: dB
R&S ZVR/ZVABT Compatible Commands