GUI Reference
R&S
®
ZNA
524
User Manual 1178.6462.02 ─ 12
ALC
Enables/disables automatic level control (ALC) for the current channel.
Same functionality as "ALC ON" on the
or in the
.
Remote command:
ALC Params
Displays the current ALC parameters. Click on the table cell to open the
that allows you to (enable and) configure automatic level control (ALC) for the
current channel.
Slope
Defines a linear factor to modify the port-specific source power as a function of the
stimulus frequency. The resulting output power at port frequency f is calculated as
<Power Result> + <Slope> * f.
Remote command:
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:SLOPe
Receiver Level Tab
Provides access to the receiver step attenuator settings.
Receiver step attenuators are optional hardware (see
Chapter 5.7.21, "Receiver Step
Attenuation
Sets the attenuation at the respective measurement receiver (b-wave). See
Remote command:
Wb/Nb IF Gain
Controls the gain of the wideband and narrowband signal paths.
Same functionality as "IF Gain Mode" on the
.
Remote command:
[SENSe<Ch>:]POWer:GAINcontrol:GLOBal
Channel Config Softtool