Example:
SOURce1:POWer:ALC:CSTAate ON
Allow automatic level control for channel 1.
SOURce1:POWer:ALC ON; SOURce1:POWer2:ALC ON
Enable automatic level control for all ports in channel 1.
SOURce1:POWer:ALC:COUPle ON
Use identical ALC settings for all ports in channel 1.
SOURce1:POWer:ALC:RANGe 5
Select a 5 dB control range (for all ALC ports).
SOURce1:POWer1:ALC:SOFFset 0
Select a start value offset of 0.
SOURce1:POWer:ALC:AUBW ON
Enable automatic bandwidth setting.
SOURce1:POWer:ALC:STOL 1
Select a 1 dB settling tolerance window.
Manual operation:
See
SOURce<Ch>:POWer:ALC:BANDwidth
<Bandwidth>
Selects the bandwidth in the ALC control loop in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Bandwidth>
Bandwidth [Hz] in the ALC control loop.
Range:
20, 30, 50, 100, 200, 500, 1E3, 1E4, 1E5, 1E6
*RST:
1E3
Default unit: Hz
Manual operation:
See
SOURce<Ch>:POWer:ALC:CSTate
<Boolean>
Enables or disables Automatic Level Control (ALC) for channel <Ch>.
In addition to the channel level, ALC can also be enabled or disabled at the port level
(
SOURce<Ch>:POWer<Pt>:ALC[:STATe] ON
).
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
ON (1)
Enable ALC for channel <Ch>
OFF (0)
Disable ALC for channel <Ch>
Manual operation:
See
SCPI Command Reference