Command Reference
R&S
®
ZNA
1143
User Manual 1178.6462.02 ─ 12
[SENSe<Ch>:]FREQuency:CONVersion
<ConversionMode>
Enables a frequency conversion measurement mode for channel
<Ch>
.
Suffix:
<Ch>
.
Channel number
Parameters:
<ConversionMode>
FUNDamental | ARBitrary | MIXer | VMIXer | IMODulation
FUNDamental
Measurement of the fundamental signal (non frequency-convert-
ing mode). This selection cancels all port-specific frequency and
power settings (see example).
ARBitrary
Frequency conversion mode (arbitrary port frequencies). This
mode is automatically activated when a port-specific frequency
is defined.
MIXer
Scalar mixer mode. The mixer mode is configured using the
[SENSe<Ch>:]FREQuency:CONVersion:MIXer...
com-
mands.
IMODulation
Intermodulation measurement mode, configured using the
[SENSe<Ch>:]FREQuency:IMODulation...
commands.
VMIXer
Vector mixer mode.
*RST:
FUNDamental
Example:
See
SOURce<Ch>:FREQuency<PhyPt>:CONVersion:
Manual operation:
See
[SENSe<Ch>:]FREQuency<Pt>:CONVersion:ARBitrary
<Numerator>,
<Denominator>, <Offset>, <SweepType>
Defines the receiver frequency formula at port <Pt> for frequency-converting measure-
ments.
For <SweepType> SWEep, the receiver frequency f
r
is calculated according to
f
r
= <Numerator>/<Denominator> * f
b
+ <Offset>,
where f
b
represents channel base frequency.
For <Sweep Type> CW and FIXed, f
r
is calculated according to
f
r
= <Offset>,
i.e. f
r
is constant.
Make sure the converted frequency or frequency range is within the analyzer's fre-
quency range.
SCPI Command Reference